Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133338

    Validate with FDA (Live)

    Date Cleared
    2013-12-20

    (51 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LigaSure™ 5 mm, Maryland Jaw, Open / Minimally Invasive Sealer/Dividers are bipolar electrosurgical instruments intended for use with the ForceTriad™ Energy Platform in general, minimally invasive and open surgical procedures where ligation and division of versels and lymphatics is desired. The instrument creates a seal by application of RF electrosirgical onergy to vascular structures (vessels and lymph) interposed between the jaws of the instrument. A blade within the instrument is surgeon actuated to divide tissue.

    Indications for use include general open and minimally invasive procedures including urologic, vascular, thoracic and thoracoscopic. and gynecologic procedures where ligation and division of the vessels is performed. These procedures include: laparoscopically assisted, vaginal hysterectomy, Nissen fundoplication, colectorny, adhesiolysis, oophorectorny, etc. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

    The LigaSure 5 mm Maryland Jaw Sealer/Dividers can be used on vessels and lymphatics up to and including 7 mm, and tissue bundles.

    Device Description

    The LigaSure™ 5 mm Maryland Jaw Sealer/Divider One-step Sealing (LF17XX series) devices are sterile, single-use, hand-held bipolar electrosurgical instruments designed for use with the ForceTriad™ energy platform (generator) to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics clamped between the jaws, grasping tissue, and blunt dissection during open and minimally invasive general surgical procedures using radio frequency (RF) energy. A hand actuated lever allows the user to open and close the instrument jaws without having to latch the lever, which includes a clicking mechanism that indicates to the user that the jaws are in the grasping zone, a button (switch) to activate the LigaSure™ mode by closing the handle against the button (switch) for vessel sealing, and a trigger to actuate an independent cutting blade.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device and outlines various tests performed to demonstrate its safety and effectiveness. However, it does not present a table of specific acceptance criteria with corresponding performance data. It also does not involve a study related to AI or human readers, as it is a physical surgical device.

    Based on the information provided, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states that evidence of safety and effectiveness was obtained from both bench and preclinical testing. While specific quantifiable acceptance criteria and their precise results are not explicitly listed in a table format, the document indicates that the device's performance was compared to predicate devices and found to be substantially equivalent.

    Here's a breakdown of the types of tests performed:

    Type of TestImplied Acceptance Criteria (Not Explicitly Stated)Device Performance (Implied from "operated as intended" and "substantially equivalent")
    Bench Testing
    IEC 60601-1 ComplianceAdherence to general requirements for basic safety and essential performance.Compliant
    IEC 60601-2-2 ComplianceAdherence to particular requirements for high frequency surgical equipment and accessories.Compliant
    Blade ReturnProper and consistent return mechanism.Satisfactory
    Grasping PerformanceSecure and effective tissue grasping.Satisfactory
    Jaw TemperatureWithin safe operating limits to prevent thermal injury.Satisfactory
    Jaw ForceAdequate force for effective sealing and grasping.Satisfactory
    Button Activation ForceOperable within ergonomic limits.Satisfactory
    Knife Deployment ForceOperable within ergonomic limits.Satisfactory
    Lever ForceOperable within ergonomic limits.Satisfactory
    Power Curve PerformanceConsistent and effective energy delivery.Satisfactory
    Preclinical Testing
    Renal and Pulmonary Burst PressureAbility to withstand physiological pressure after sealing.Effective sealing up to and including 7 mm vessels.
    Sealing of Vessels & Lymphatics (up to 7mm)Effective and durable seal formation.Effective sealing.
    Hemostasis of Tissue & VesselsAbility to stop bleeding effectively.Achieved hemostasis.
    Tissue BundlesEffective sealing and division of tissue bundles.Effective sealing and division.
    Thermal SpreadMinimal spread of thermal energy to surrounding tissue.Minimal and acceptable.
    Lymphatic Burst PressureAbility to withstand physiological pressure after sealing.Effective sealing.
    Chronic Animal StudyLong-term safety and effectiveness in vivo.Satisfactory, results support substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state specific sample sizes for each test. It broadly mentions "bench and preclinical testing." It also does not specify the country of origin for the data or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. Given that this is a physical electrosurgical device for which "ground truth" would be established through physical measurements, physiological responses, and potentially expert observation during preclinical studies (e.g., assessing hemostasis), the concept of "experts establishing ground truth for a test set" as described in the prompt (e.g., radiologists) is not directly applicable in the same way it would be for an AI-powered diagnostic device.

    4. Adjudication Method for the Test Set

    This information is not provided and would not typically be relevant for the type of device and testing described. Adjudication methods like 2+1 or 3+1 are more commonly used in studies involving human interpretation of data, such as medical image analysis.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. without AI Assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI algorithms on human performance in diagnostic tasks, which is not applicable to this physical electrosurgical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    This question is not applicable. The device is a physical electrosurgical instrument and does not involve an "algorithm only" component in the sense of a software-based diagnostic tool. The performance studies described are for the physical device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance criteria appears to be established through:

    • Physical measurements and engineering standards: For bench tests like blade return, jaw temperature, forces, and power curves. These would be compared against predefined specifications and industry standards (e.g., IEC 60601).
    • Physiological outcomes in preclinical (animal) studies: For burst pressure, hemostasis, thermal spread, and the success of vessel/tissue bundle sealing and division. These outcomes would be assessed by observing the effects on biological tissue.

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable. The device development process described does not involve a "training set" in the context of machine learning (AI). The device's design and engineering are based on established principles and iterative development, not data-driven machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided and is not applicable for the reasons stated above (no AI training set).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1